Texture image
Exradin a600 icon 1400

Diagnostic

Exradin® Magna A600 Ion Chamber, Diagnostic Parallel Plate Chamber, 1.50cc

The A600 chamber is designed for consistent air kerma, absorbed dose and exposure measurements. Vented and fully-guarded, this chamber is perfectly suited for mammography and general diagnostic x-ray regions.

Exradin® Magna A600 Ion Chamber, Diagnostic Parallel Plate Chamber, 1.50cc part of complete solution for: Diagnostic CT QA

Active Volume

  • A600 Chamber
    1 cm3
  • A650 Chamber: 3 cm3

Materials

  • Air equivalent plastic, acrylic

Entrance window

  • Kapton conductive film

Connector

  • Two lug triax (standard)
  • Center pin
    Collector
  • Inner Guard
    Chamber Guard
  • Outer Housing
    HV GND

Cable Length

  • 1 m (39.4 inches)

Bias Voltage

  • Normal operation at +/-300 Vdc

Nominal Inherent-Leakage Current

  • 10-15 A

Reference Point

  • Scribed line around circumference of the outer ring is halfway between the collector and the entry window.

Accuracy

  • +/- 1.0 %

Stability

  • 0.2%

Energy Response

  • 92600
    from 30kVp to 150kVp + 2%
  • 92650
    from 30kVp to 150kVp + 4%

Nominal Calibration Factor

  • A600
    3 R / nC
  • A650
    9 R / nC

Aion (typical)

  • 1.000

Chamber Dimensions 92600

  • Height
    2.54 cm
  • Outside Diameter
    4.13 cm
  • Inner Plate separation
    8 mm
  • Weight
    0.2 kg

Chamber Dimensions 92650

  • Height
    2.08 cm
  • Outside Diameter
    5.33 cm
  • Inner Plate Separation
    8 mm
  • Weight
    0.2 kg