SuperMAX Electrometer

The premier two-channel, reference-grade electrometer on the market.

SuperMAX™ Electrometer part of complete solution for: Stereotactic Absolute Dosimetry CyberKnife

SuperMAX™ Electrometer Overview

Two Independent Measurement Channels

  • Independent control over range, bias voltage, and applied system factor.
  • Both channels feature extensive range (0.001 pA to 500.0 nA, 0.001 pC to 999.9 µC) for use in a variety of applications, including ratio measurements.

Built-In Detector Library

  • Eliminates the need for a PC or to apply system factor and temperature/pressure corrections.
  • Store over 100 detectors and calibrations.
  • Apply individual factors to one or both measurement channels.

Easy-to-Use Interface

  • Large, color display with built-in touchscreen.
  • On-screen keypad and pull down menus allows for easy access to settings and detector library.

Store Data Collection

  • Measurements are stored from all data collection.
  • Set automatic repeat measurements.

Exradin W1 Scintillator Integration

Two dedicated modes in the SuperMAX allow for seamless calculations using the Exradin W1 Scintillator.

Rate Mode Display Ranges

  • LOW: 0.001 pA – 500.0 pA
  • HIGH: 0.001 nA – 500.0 nA

Charge Mode Display Ranges

  • LOW: 0.001 pC – 999.9 μC
  • HIGH: 0.001 nC - 999.9 μC

Rated Effective Range on Input Currents Per IEC 60731 for Reference Grade

  • 0.400 pA – 500.0 nA (± .25% of minimum effective scale reading)

Measuring Range

  • LOW: 0.400 pA - 500 pA, 0.400 pC - 999.9 μC
  • HIGH: 0.400 nA - 500 nA, 0.400 nC - 999.9 μC

Bias Voltage

  • Nominal ±1000 V
  • 0, (±) 100 - 1000 V in 1 V increments

Range Switching

  • User selectable

Display Resolution

  • LOW: Rate mode/Charge mode
    0.001 pA / 0.001 pC
  • HIGH: Rate mode/Charge mode
    0.001 nA / 0.001 nC

Repeatability

  • Low range: 0.25%
  • High range: ± 0.1%

Long Term Stability

  • ± 0.5% over 1 year

Stabilization Time

  • ± 0.1% of value at 1 hour for measurements taken at 15 minutes and 6 hours

Non-linearity:

  • ± 0.25%

Zero Drift

  • Rate mode: ± 0.25% of minimum effective scale reading (LOW/HIGH: < ± 0.001 pA / < ± 0.001 nA)

Zero Shift

  • Rate mode: ± 0.25% of minimum effective scale reading (LOW/HIGH: < ± 0.001 pA / < ± 0.001 nA)

Charge Leakage Current

  • LOW/HIGH: < ± 0.001 pA / < ± 0.001 nA

Response Time

  • Low range rate: < 12 seconds, High range rate: < 2 seconds
  • Low range charge: <2 seconds, High range charge: <0.3 seconds

Main Software Functions

  • Channel 1 Measurement, Channel 2 Measurement, Channel 1 & 2 Measurement, Scintillator Measurement, Check Source Library, Chamber Library, Scintillator Baseline Library, Preferences

Charge Collection Modes

  • Timed Mode - Selectable timer from 1 to 600 seconds in 1 second increments
  • Continuous Mode (Manual Start/Stop) - Press Start to begin collection and Stop to end
  • Trigger Mode – current exceeding pre-selected trigger values starts and stops collection

Countup/Countdown Resolution

  • 1 second (count up for Continuous Mode, count down for Timed Mode)

Trigger Levels for Charge

  • Collection
    Start value must be larger than stop value (set via Preferences interface)
  • Start
    0.2 - 9.9 pA
  • Stop
    0.1 - 9.8 pA

Operating Parameters

  • Temperature
    15 to 35 °C
  • Relative Humidity
    20 to 80% non-condensing
  • Pressure
    650 to 770 mmHg

Storage Parameters

  • Temperature
    -15 to 50 °C
  • Relative Humidity
    10 to 95% non-condensing
  • Pressure
    600 to 800 mmHg

Display

  • Color LCD TFT touch screen. VGA (640x480) resolution, 6.4” diagonal

Input

  • 2 channels, BNC two lug triaxial connector (TNC option is available.)

Bias Voltage

  • ± VDC for the following: 0 to 1000 VDC in 1 VDC increments, 350 μA max

Zeroing

  • Zero via button push - Display indicates zeroing in progress

Ports

  • RS232 (Default: Uni-directional 19,200 baud, 8 data bits, no parity, 1 stop bit; Bi-directional operation via user commands, tech note available upon request) (2) USB ports (keyboard, thumb drive, or mouse support) Ethernet (internal use only, not currently supported)

Chamber library calibration coefficients

  • Amp System factors
    Nsk:μGym²/(h*A)
  • Nsk:Gym²/(h*A)
  • Nx,sk:Rm²/(h*A)
  • NA:GBq/A
  • NA:Ci/A
  • ND,W:mGy/(sec*A)
  • NA:MBq/A
  • NA:mCi/A
  • Coulombs System Factors
    Nk:Gy/C
  • Nx:R/C
  • NGas:Gy/C
  • ND:Gy/C
  • ND,W:Gy/C

Real-Time Clock

  • On board, for data validation and storage

Audio

  • On board speaker for audio feedback on start-up, button presses, operator alerts, etc.

Dimensions

  • Width
    26.7 cm (10.5 in)
  • Height
    8.1 cm (3.2 in)
  • Length
    21.1 cm (8.3 in)
  • Weight
    2.5 kg (5.45 lbs)
  • Mode of Operation
    Continuous

Power Requirements

  • 100-240 VAC, 0.5 A max, 50/60 Hz input to external power supply, 9 VDC, 2.0 A power supply output to electrometer input, Globtek®, Inc. model GTM 96180-1811-2.0-T.
  • The use of any other power supply and using alternates other than the UL/CSA recognized power cord can degrade minimum safety. The proper replacements from Standard Imaging, Inc. are required for compliance with the requirements of IEC 60601-1.

Product Standards

  • IEC 60601-1:3rd Edition1, IEC 607312
  • 1 Externally Certified; 2 Designed to Meet

Shock Classifications

  • Class I - External Power Supply
  • Class II - Electrometer

Videos

Tech Notes

Please sign in or register to the customer care portal to view tech notes.

Have Questions About the SuperMAX™ Electrometer?

This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.

Request Pricing for Complete CyberKnife Solution

Filmless and finished in under 10 minutes. Receive more information about CyberKnife QA Solutions.

Learn more about Standard Imaging CyberKnife

Join our Mailing List